ZHAN Zhao, ZHANG Yan, GUI Xing-min. Experiment and design improvement of an axial flow blood pump[J]. Journal of Aerospace Power, 2008, 23(6): 1061-1066.
Citation: ZHAN Zhao, ZHANG Yan, GUI Xing-min. Experiment and design improvement of an axial flow blood pump[J]. Journal of Aerospace Power, 2008, 23(6): 1061-1066.

Experiment and design improvement of an axial flow blood pump

  • Received Date: 2007-05-24
  • Rev Recd Date: 2007-06-25
  • Publish Date: 2008-06-28
  • The numerical investigation and experimental results of the proto-pump LAP-00 show that at design rotational speed of 9000 r/min,the flow rate corresponding to the highest efficiency point is 11 L/min,which is much higher than clinical use(38 L/min),causing the pump to work at off design point in clinic use with a low efficiency.Besides,in the diffuser region,flow separation and recirculation may cause blood stagnation and thrombosis.The improved design for pump LAP-23 decreases the design rotational speed from 9000 r/min to 8000 r/min,thus reduces the high shear stress on blood cells at impeller tip;meanwhile,it decreases the design flow rate to 6 L/min and eliminates flow separation and recirculation in the diffuser region.

     

  • loading
  • [1]
    云忠,龚中良.旋转叶轮血泵的发展与展望[J].生物医学工程学杂志,2005,22(1):151-154.YUN Zhong,GONG Zhongliang.Development and expectation of rotary impeller blood pump[J].J Biomed Eng,2005,22(1):151-154.
    [2]
    Greg W B.James F A,Wu Z J,et al.Computational fluid dynamics as a development tool for rotary blood pump[J].Artificial Organs,2001,25(5):336-340.
    [3]
    王芳群,曾培,封志刚,等.应用CFD研究叶轮设计对人工心脏泵内流场的影响[J].中国生物医学工程学报,2005,24(5):578-582.WANG Fangqun,ZENG Pei,FENG Zhigang,et al.An investigation of effect of impeller design on flow patterns in blood pumps with CFD[J].Chinese J Biomed Eng.2005,24(5):578-582.
    [4]
    Kerkhoffs W,Schumacher O,Meyns B.Design,development,and first in vivo results of an implantable ventricular assist device,MicroVad[J].Artificial Organs,2004,28(10):904-910.
    [5]
    Apel J.Paul R.Klaus S,et al Assessment of hemolysis related quantities in a microaxial blood pump by computational fluid dynamics[J].Artificial Organs,2001,25(5):341-347.
    [6]
    薛嵩,张岩,桂幸民.运用三维数值模拟对人工心脏血液泵的设计和改进[c]//北航第二届研究生学术论坛文集.北京:北京航空航天大学出版社,2005.XUE Song,ZHANG Yan,GUI Xingmin.Digital stimulation to development of axial pump for artificial heart[C]//Collected works of the 2nd Postgraduates' Forum of BUAA.Beijing:BUAA Press,2005.
    [7]
    Huang C R,Fabisiak W.A rheological equation characterizing both the time dependent and the steady state viscosity of whole human blood[R].AICHE Symposium Series,1978:19-21.
    [8]
    Wells R E,Merrill E W.Shear rate dependence of the viscosity of whole blood and plasma[J].Science,1961(133):763-764.
    [9]
    Untaroiu A,Wood H G.,Allaire P E,et al.Computational design and experimental performance testing of an axial flow pediatric ventricular assist device[J].ASAIO Journal 2005,(51):629-635.
    [10]
    Tetsuya Yano,Kazumitsu Sekine,Ayumi Mitoh,et al.An estimation method of hemolysis within an axial flow blood pump by computational fluid dynamics analysis[J].Artificial Organs,2003,27(10):920-925.
    [11]
    Apel J,Neudel F,Reul H.Computational fluid dynamics and experimental validation of a microaxial blood pump[J].ASAlO,2001.(47):552-558.
    [12]
    Giersiepen M.Wurzinger L J,Opitz R,et al.Estimation of shear stress-related blood damage in heart valve prostheses-in vitro comparison of 25 aortic valves[J].Artif Organs,1990,13:300-306.
    [13]
    Bludszuweit C.Three-dimensional numerical prediction of stress loading of blood particles in a centrifugal pump[J].Artif Organs,1995,19:590-596.
    [14]
    Paul R,Apel J,Klaus S,et al.Shear stress related blood damage in laminar couette flow[J].Artif Organs,2003.27:517-529.
  • 加载中

Catalog

    通讯作者: 陈斌, bchen63@163.com
    • 1. 

      沈阳化工大学材料科学与工程学院 沈阳 110142

    1. 本站搜索
    2. 百度学术搜索
    3. 万方数据库搜索
    4. CNKI搜索

    Article Metrics

    Article views (1794) PDF downloads(456) Cited by()
    Proportional views
    Related

    /

    DownLoad:  Full-Size Img  PowerPoint
    Return
    Return